Switchback 6600 CE Power Supply/Temperature Controller

Switchback Power Supply by Solid State Cooling SystemsThis bipolar current-controlled power supply has been optimized for use with our thermoelectric heat exchangers

Current Customer Applications

  • Paired with Cleanstream or Terraview thermoelectric heat exchangers to control temperature of a wide range of aqueous, ultra-high purity or corrosive solutions and liquid chemicals, including Fluorinert, Galden and HFE
  • Part of custom air-to-liquid systems for semiconductor and military applications
  • Semiconductor processing

Download PDF Brochure

Overview

Why choose the Switchback Power Supply/Temperature Controller?

  • This power supply was created to provide a complete package for the operation of the Cleanstream and Terraview heat exchangers.
  • Precise temperature control (±0.03 °C) and excellent energy efficiency.
  • Reliable, ripple-free, continuously-variable DC power to cool and heat a thermoelectric heat exchanger
  • 16 bit PID + fuzzy logic temperature control
  • Variable output current ranges from 0-6 to 0-30 amps at voltages up to 230 VDC
  • Integrated phase monitor and current rate limiter for reliable thermoelectric lifetime.

Features

Communication Options:
Option 1: Remote 0-10 VDC input for set- point temperature
Option 2: RS-485 communication + remote input

Alarms: Four programmable temperature alarms

Digital Inputs: Three digital inputs for selecting pre-programmed temperature setpoints

Analog Outputs: 4-20 mA retransmission of present temperature value standard

Interlocks:
Three interlocks for automatic shut down of output current:
RTD fault
Power supply fault
Thermostat fault

Specifications

Specifications Summary – Switchback Power Supply/Temperature Controller

Temperature Sensor Required: 3-wire 100 W Pt RTD, α =.0385
Input Power: 3-phase 208 VAC, 20 amps (some applications require 30 amps)
Output Power: Bipolar, 0-230 VDC, 0-30 amps, user-adjustable maximum output ranges from 14-230 VDC and 6-30 amps.
Dimensions: 19 x 5.25 x 17” (19” rack-mount), 48.3 cm wide by 13.3 cm tall x 43.2 cm deep
Weight: 24 lbs. (10.9 kg)
2 Year Warranty

Switchback 6600 CE Power Supply Specifications

Options Purpose
Input Power: 200-240 VAC (±5%), 3-phase 22 amps, 50 or 60 Hz, 3-wire plus ground
Line Filtering: Integral line filter meets CE conducted emissions requirement
Rated Capacity: 6.9 kW
Output Current: 0-30 amps
Output Voltage: 0-230 VDC
Output Current Control Range: Maximum output current (IMAX) is dipswitch adjustable from 6 to 30A.
Output Voltage Limit: Maximum output voltage (VMAX) is dipswitch adjustable from 14 to 230 VDC.
Non-isolated Output: The power supply output is not isolated from the AC input. The output must not be connected to secondary referenced circuitry. Input voltage must be disconnected to prevent hazardous potentials from appearing at the output. DC control and logic circuits are isolated from the AC line and DC output.
Voltage Ripple: <2.5 volts for 0-50 VDC; <5% for 50-220 VDC
Regulation: <5% of actual output current
Control Band Width: The current control loop bandwidth varies with the load resistance (RLOAD) by the following relationship: Bandwidth = 5288/RLOAD (HZ)
Transient Response: Output current will regulate within 10 milliseconds of a setpoint change or load transient of up to 50%
Operating Temperature: 0-40 °C, no derating
Operating Humidity: 0-80% relative humidity, non-condensing
Cooling: External 4.69" fan (provided). Leave >1" clearance around air intake and exhaust.
Control Signals: Output current proportional to 0-5 VDC input signal (0 VDC = 0 amps, 5 VDC = IMAX); current polarity reversed via TTL signal, On/Off via TTL signal.
Optional Control Signals: Dual 4-20 mA. Normal polarity (cooling) output current proportional to input 1, reverse polarity (heating) output current proportional to input 2. 4 mA = 0 amps, 20 mA = IMAX
Fault Signal: A TTL high signal indicates normal operation. If a thermal shutdown or output stage overload occurs, a TTL signal low is produced.
Local/Remote Operation: Output current and polarity may be controlled via one of two sources: "remote" or "local"
Remote Sensing: A 0-5 VDC signal proportional to the output current at Imax/5 amps per volt, where Imax is the maximum output current setting
Over-current Protection: Automatic electronic current limiting is provided
Thermal Protection: Automatic shutdown when heat sink reaches 70 °C; automatic restart at 53 °C
Isolation: Can withstand a 3000 VDC Hipot with input and output terminals tied together
Courtesy Outputs: 12 VDC, 150 milliamps; 5 VDC, 20 milliamps

End Slides

Share this product